StutzendurchmesserDN50/2"
Maximale Stutzenlänge [mm (in)]50 (1.97) mm (in)
Masaca. 0,9 kg (1.98 lbs) ohne Überflutungsschutzhülse
Abmessungen der Ausführung 583 mm583 mm (23 in) mm (in)
Abmessungen der Ausführung 1 085 mm1 085 mm (42,7 in) mm (in)
Materiał316Ti (1.4571)
Prozesstemperatur−40...+80 °C(−40...+176 °F)
Prozessdruck0,7...4 bar (10.15...58 psi)
KabeltypLiYCY/CUL 2x(0,75)
Gewicht (einschließlich 5 m (16 ft) Kabel)ca. 0,9 kg (1.98 lbs) ohne Überflutungsschutzhülse
Blockdistanz (BD)0,07 (0.2) m
24 VDC± 10%
Restwelligkeit< 100 mV
max. Reichweite in Flüssigkeiten3m
max. Reichweite in Schüttgütern1,2m
blocking distance (BD)0.07 m
Max. range1 in solids1.2 m
Tension d’émission55 Veff
Fréquence d’émission (20 °C)90,0 kHz
Consommation maximale0,9 W
Température de process maximale+60 °C
Gamme de température ambiante admissible–40 °C...+60 °C
Température de surface maximale80 °C
max. Prozesstemperatur+60 °C
Zone 2 - ApplicationFDU90-E..A., FDU91-E..A.
Température de surface maximale pour température ambiante maximale de 40 °C80 °C
Température de surface maximale pour température ambiante maximale de Tmax *100 °C
Max. range in solids1.2m
Circuit du capteur :FMU90, FMU95 à FDU9x
Blockdistanz7cm
Maximum measuring distance1)3 m
Maximum measuring distance3 m
Max. range in liquids3m
Process temperature–40 to +80 °C (–40 to +176 °F)
Maximum measuring range in liquids3 m (9.8 ft)
Current consumption250 mA per sensor
Operating frequency90 kHz
Process pressure0.7 to 4 bar (10.15 to 58 psi)
Maximum measuring range3 m (9.8 ft) in liquids; 1.2 m (3.9 ft) in bulk solids
Attenuation caused by surface of liquid - Waves on surface5 to 10 dB
Attenuation caused by surface of liquid - Very turbulent surface10 to 20 dB
Attenuation due to bulk solids surface - Hard, rough surface40 dB
Attenuation due to bulk solids surface - Soft surface40 to 60 dB
Attenuation due to dust - Minor dust formation5 dB
Attenuation due to dust - Major dust formation5 to 20 dB
Attenuation caused by filling curtain in detection range - Small volumes5 dB
Attenuation caused by filling curtain in detection range - Large volumes5 to 20 dB
Attenuation caused by temperature difference between sensor and product surface - Up to 40 °C5 to 10 dB
Attenuation caused by temperature difference between sensor and product surface - Up to 80 °C10 to 20 dB
Weight excluding flooding protection tubeapprox. 0.9 kg
Weight including flooding protection tubeapprox. 1.0 kg
Weight including cable 5 m (16 ft)approx. 1.0 kg (2.21 lb)
Α (typically)12°
(typically)12°
Maximum measuring distance1)3,0 (9.8) (for liquids)