OceanSphereENGINES

FMP52

Radar Level Gauge · Cargo Level & Temperature Gauging

Class-certified type (DNV register) · technical data sheet not yet verified​⁠​‌​‍⁠‍​‍‌​‍​‍⁠‍‍‌​
Specifications

Technical data.reference — unverified

Do you have the manufacturer datasheet for this model? Send it with the enquiry form — we check it and name the source.

Position of 1st measuring pointA = 350 mm (13.8 in)
Minimum measuring rangeB ≥ 400 mm (15.7 in)
Minimum length of probeLN ≥ 1 000 mm (39.4 in)
Terminal voltage U at device13.5 to 30 V
Maximum load R, depending on the supply voltage U0 of the power supply unit500
Power supply; output90 to 253 VAC (50 to 60 Hz), overvoltage category II
Maximum load Rmax500 Ω
Terminal voltage U9 to 32 V
Power consumption< 0.9 W
Nominal current HART3.6 to 22 mA
Breakdown signal HARTadjustable: 3.59 to 22.5 mA
Nominal current PROFIBUS PA14 mA
Probe length< 10 m (33 ft)
Sampling rate3.6 measurements per second
Response time< 0.8 s
Rod 16 mm (0.63 in) PFA>316L30 Nm
Flange size10K 50A
Tightening torque40 to 60 Nm
Process temperature–50 to +200 °C (–58 to +392 °F)
B78 (3.1) mm
C78 (3.1) mm
D78 (3.1) mm
I10 mm
J14 mm
K39 mm
L1000 (39.4) mm
M500 (19.7) mm
N40 (1.57) mm
Position of 1st measur-ing pointA = 350 mm (13.8 in)
Position of 3rd measur-ing pointC = 250 mm (9.84 in)
Minimum probe lengthLN ≥ 1 000 mm (39.4 in)
Minimum distance between reference point R and 100% markA ≥ 250 mm (10 in)
Rope probe ... mm (... in), 4 mm (¹⁄₆ in)PFA > 316, max300 mm (12 in)nozzle height, centering rod
Terminal voltage U at the device13.5 to 30 V
Maximum load R, depending on the supply voltage U0 at the supply unit500
Power supply; Output9 to 32 V 3)
Temperature derating for FMP54 - version HT+450 °C (+842 °F )
Ta+80(+176)
Gt18+54(+129)
Tp-40/-50 (-40/-58)³
Gt20+53(+127)
Gt19+38(+100)
Ln1000 (39.4) mm
Rope probe 4 mmApprox. 0.5 kg/m probe length
Rod probe 16 mmApprox. 1.1 kg/m probe length
Minimum distance from the probe end to the 0% markC ≥ 100 mm (4 in)
Position of 3rd measuring pointC = 250 mm (9.84 in)
Rod 16 mm (0.63 in), PFA over 316L30 Nm
A4...20 mA HART
Process temperature range–50 to +200 °C (–58 to +392 °F)
Probe typeRope probe ... mm (... in), 4 mm (¹⁄₆ in)PFA over 316, max. 300 mm (12 in) nozzl
Fw18
Dev.rev.16
Bending strength of rod probes [Nm]30 Nm
Max. nozzle height (= length of the center rod)150 mm
Bending strength [Nm]30
Øb [mm (inch)]< 14 (0.55) mm (inch)
Ø<25 (1.0) [mm (inch)]8.5 (0.34) mm (inch)
»3 (0.12) [mm (inch)]12.5 (0.52) mm (inch)
»50 (1.97) [mm (inch)]16.5 (0.65) mm (inch)
Minimum bending radius100 mm (4 inch)
Max. nozzle height150 mm
Recommended torque [Nm]35 minimum
Rmin100 (4)
Rope4 mm (¹⁄₆ in) 316
Tensile loading capacity10 kN
Maximum wall thickness (a):Plastic: < 15 mm (0.6 in), Glass: < 10 mm (0.4 in)
Resistance per channel:2 × 0.5 Ωmax
Threshold DC voltage:400 to 700 V
Threshold surge voltage:< 800 V
Capacitance at 1 MHz:< 1.5 pF
Nominal leakage current (8/20 μs):10 kA
Suitable for conductor cross-sections:0.2 to 2.5 mm2 (24 to 14 AWG)
Cable specificationScrew terminals for wire cross-sections 0.2 to 2.5 mm2 (24 to 14 AWG)
For ambient temperature TU≥60 °C (140 °F):use cable for temperature TU +20 K
Max. probe length10 m (394 in)
Bending strength of rod probes30 Nm
Permissible nozzle height≤ 150 mm (6 in)
Seitliche Belastbarkeit (Biegefestigkeit) von Stabsonden30 Nm
Stab 16mm (0,63') PFA>316L30 Nm
Einsatzhöhe nach IEC61010-1 Ed.3bis 2 000 m (6 600 ft) über Normalnull
Prozesstempertaur–50 … +200 °C (–58 … +392 °F)
Sensorca. 1,2 kg + Flanschgewicht
Seilsonde 4 mmca. 0,5 kg/m Sondenlänge
Stabsonde 16 mmca. 1,1 kg/m Sondenlänge
ZylinderschraubeA4-70 + Scheibe; A4
Ad2000W2 und W10
Mindestabstand zwischen Referenzpunkt R und 100%-MarkeA ≥ 250 mm (10 in)
Minimaler MessbereichB ≥ 400 mm (16 in)
Mindestabstand vom Sondenende zur 0%-MarkeC ≥ 100 mm (4 in)
Maximalwert für "Abgleich Leer"E ≤ 3,9 m (12,8 ft)
SchutzartIP68 / NEMA 6P und IP66 / NEMA 4x
Passend für die AnzeigemoduleSD02 (Drucktasten), SD03 (Touch control)
VerbindungskabelMitgeliefertes Kabel bis 30 m (98 ft), Kundenseitiges Standardkabel bis 60 m (19
Umgebungstemperatur–40 … 80 °C (–40 … 176 °F)
Optionale Umgebungstemperatur–50 … 80 °C (–58 … 176 °F)
Ex-frei11,5 … 35 V
Ex ic500
Ex ia / IS11,5 … 30 V
Ex d / XP500
Klemmenspannung U am Gerät13,5 … 30 V 3) 4)
Maximale Bürde R, abhängig von der Versorgungsspannung U0 des Speisegeräts500
Genauigkeit (Summe aus Nichtlinearität, Nichtwiederholbarkeit und Hysterese)Füllstandmessung: ±0,02 %
Nichtwiederholbarkeit≤ 1 mm (0,04 in)
Sondenlänge1 000 mm (40 in)
Seitliche Belastung (Biegemoment) durch StrömungM = cw × ρ/2 × v2 × d × L × (LN - 0.5 × L)
Max40 Prozessanschluss mit Flansch -
Temperatur-derating-40(-40) °C
CaStab 16mm
CbStab 0,63in
OaSeil 4mm, max. 150mm Stutzen
ObSeil 4mm, max. 300mm Stutzen
2Beschichtung 2 mm (0,08 in): PFA
Seil4 mm (¹⁄₆ in) PFA>316
Stab 16 mmPFA>316L 30 Nm
Flanschgröße10K 50A
Anzugsdrehmoment40 … 60 Nm
Werkstoff mediumberührtPFA, PTFE (FDA, 3A, EHEDG, USP Cl. VI zugelassen) –50 … +200 °C (–58 … +392 °F)
Ø1616 mm
Ø16 (0.63)0.63 in
Ø50.5 (1.99)50.5 mm
Ø37 (1.46)37 mm
Ø22 (0.87)22 mm
-196(-321) °C ([ °F])
-40(-40) °C ([ °F])
+81+178 °C ([ °F])
+450+842 °C ([ °F])
Material2 ASME: 316/316L
drucktragende metallische Werkstoffeerfüllen die Anforderungen der NACE MR 0175 / ISO 15156
Konformitätserklärungsiehe Produktstruktur, Merkmal 580, Ausprägung JB
Test, ZeugnisKonformitätserklärung NACE MR0175, mediumberührte metallische Teile
Position 1. MesspunktA = 350 mm (13,8 in)
Position 3. MesspunktC = 250 mm (9,84 in)
Minimaler Messbereich B≥ 400 mm (15,7 in)
Minimale Sondenlänge≥ 1 000 mm (39,4 in)
Seilsonde,,, mm (,,, in), 4 mm (¹⁄₆ in)PFA>316, max 300 mm (12 in)Stutzenhöhe, Zentrierst
Seilsonde 4 mm oder ¹⁄₆"ca. 0,50 kg/m (1,10 lb/in)
K/lXA01041F
Seitliche Belastbarkeit30 Nm
⌀12 mm (0,47 in)
Seitliche Belastbarkeit (Biegefestigkeit) [Nm]30 Nm
Max. Stutzenhöhe150 mm
Trennschicht Kriterium1184 mV
Gemessene Kapazität1066 pF
Blockdistanz0 ... 200 m
Rampe bei Echoverlust0,0 %/min
empfohlenes Anzugsdrehmoment [Nm]35 Nm
Stutzendurchmesser≥ 400 mm (16")
Max. Temperature280 °C (536 °F)
ApprovalNEPSI Ex ia IIC T6 Ga/Gb, Ex iaD 20/21 T85…90°C
Versorgungsspannung2-Draht; 4-20mA HART, passiv
A [mm (inch)]< 20 (0,78)
B [mm (inch)]12,5 (0,52)
Temperatur-50 bis +200 °C
Druckbis 40 bar
Measuring rangeStab bis 4 m, Seil bis 45 m
Failure current FDE (Fault Disconnection Electronic)0 mA
Failure current FDE PROFIBUS PA0 mA
Rope 4 mm (¹⁄₆ in) PFA>3162 kN
Manufacturer documentation

Images from manufacturer documentation.

Photos, drawings and page previews from this model’s datasheet or catalog.

Levelflex FMP52 and FMP54 are level sensors used for measuring liquid levels in tanks and vessels.Source: Endress+Hauser FMP52 datasheet, p. 77

Applications

  • level measurement

Features

  • Tri-Clamp connections

Options

  • optional with centering star
Source: Endress+Hauser FMP52 datasheet, p. 69
Endress+Hauser FMP52 – Drawing (3ded0d936c1a_S1-127 S. 1)
Drawing · 3ded0d936c1a_S1-127 · S. 1
Endress+Hauser FMP52 – Page preview (3ded0d936c1a_S1-127 S. 5)
Page preview · 3ded0d936c1a_S1-127 · S. 5
Endress+Hauser FMP52 – Page preview (3ded0d936c1a_S1-127 S. 6)
Page preview · 3ded0d936c1a_S1-127 · S. 6
Endress+Hauser FMP52 – Page preview (3ded0d936c1a_S1-127 S. 7)
Page preview · 3ded0d936c1a_S1-127 · S. 7
Source: Endress+Hauser FMP52 datasheet, p. 82
Need this part?

Request a quote for this model.

Free of charge — send us your request for Endress+Hauser equipment. Our team checks it and replies by e-mail.

For marine equipment hardly any maker or authorised dealer will name a price while the vessel is unknown. With the vessel details, availability and price can be settled faster; without them we may have to ask first, which costs a few days. If you are not allowed to name the vessel, a photo of the rating plate is enough — attach it below or send it as a reply to our confirmation e-mail.

Your contact details are not published. No account required.
Related models

Also in Radar Level Gauge.

Other models of the same equipment type, Endress+Hauser first.

By manufacturer Endress+HauserKrohneEmerson RosemountVEGATokyo KeikiBarksdaleDanfossPanasiaSiemensNORIS AutomationEmersonHanla IMSHoneywellIfm ElectronicJUMOAll manufacturersSpare parts by equipmentNew here? Create a free account Can't read the plate? Try the nameplate scanner All Endress+Hauser models More like FMP52
More than the data sheet shows

We answer by e-mail, never with an automatic reply.

Tell us what you need for this Endress+Hauser FMP52 — a part, a service call, a second opinion. Are you Endress+Hauser? Claim this brand page.

Service engineer? Write the day report for this FMP52 in OceanSphere Work — photos, hours and parts, as a PDF for the owner. Try it free for 14 days →

Frequently asked

Questions that come up with this equipment.

Is this model's data free?

Yes. Specifications, documents and service notes for this model are readable without an account. An account adds saved searches and requests, nothing is taken away.

Can you find a part for this exact model?

Yes. Tell us the manufacturer, the model and what you need — a person reads every request and replies by e-mail.

The specifications look wrong — what now?

Send us the nameplate or the correct value. We check it against the manufacturer datasheet or class-society record and update the page.