PAH Sensor enviroFlu-HC 500, enviroFlu-HC 5000, PAH Sensor enviroFlu-HC MB 500, enviroFlu-HC MB 5000
PAH Sensor enviroFlu-HC 500, enviroFlu-HC 5000, PAH Sensor enviroFlu-HC MB 500, enviroFlu-HC MB 5000 by TriOS Mess- und Datentechnik GmbH. DNV TA certificate TAA00002FB.
Inspector Detail
▸
Common issues
- Optical window fouling/biofouling reducing sensor sensitivity and causing false low readings
- Calibration drift over time requiring periodic re-zeroing against a reference standard
- Air bubbles in the flow cell causing erratic or noisy readings
- Cable/connector corrosion at wet-space connections
- LED/light source degradation over long service life reducing signal strength
- Flow cell blockage from debris in the sample stream
Inspection checklist
- Inspect the optical window for fouling, scratching, or biofilm and clean per maker procedure
- Verify calibration against a known reference/check standard
- Check sample flow rate through the sensor's flow cell is within specification
- Inspect cable and connector for damage or corrosion
- Confirm sensor output correlates sensibly with recent discharge records
- Check power supply/signal cable integrity to the control/alarm unit
- Verify alarm setpoints match the applicable discharge limit for the equipment served
- Review the sensor's self-diagnostic/error log if provided
Service intervals
| Optical window cleaning | monthly to quarterly depending on fouling rate, more often in biologically active water |
| Calibration check | every 12 months, as part of the associated ODME/OWS certification |
| Full sensor service by maker or approved agent | per the type approval certificate interval |
| Flow cell blockage inspection | at each maintenance round |
Critical spares
- Optical window/lens assembly
- Sensor cable/connector assembly
- Calibration reference standard/check solution
- O-rings/seals for the flow cell housing
- Complete replacement sensor head/probe, the typical service approach for this class of optical sensor when out of tolerance
Safety
- Normal hydrocarbon handling precautions apply, as the sensor is typically installed in an oily bilge water/hydrocarbon system
- Electrical hazard if serviced without isolating power
- A fouled or miscalibrated sensor can let an illegal discharge go undetected - correct function is environmentally as well as physically safety-critical
Class survey
Where fitted as part of Oil Discharge Monitoring Equipment (ODME) or an oily water separator 15 ppm alarm system, this sensor's type approval certificate and calibration record must be current and match the equipment fitted. Surveyors typically check the calibration log, alarm setpoint configuration, and that the unit is included in the vessel's IOPP Certificate supplement.
Components & Design
Component data being added…
Technical Data
🤖 Ask about this model
One free question — answered like a marine engineer. No account needed.
Suppliers & Spare Parts
Direct links to manufacturer, datasheet, spare-parts portal and dealers — request access below.
📄 Request sourcing links & documentsCases, daily reports, timesheets and AI troubleshooting in the OceanSphere Marine app — free for 14 days.
Start free trial →Find manufacturers, repair shops and spare-part dealers in the Maritime Network — our global company directory.
Maritime Network →DNV type approval
Verified against the DNV Approval Finder — the certificate below is issued by DNV, not by us.
- Certificate
- TAA00002FB
- Approved as
- Miscellaneous Transmitter · Instrumentation and Automation
- Certificate holder
- TriOS Mess- und Datentechnik GmbH (Germany)
- Valid until
- 2029-02-04